MD-480A Phaco Emulsifier
Features
* All parts of the system are fully controlled and monitored by Microprocessor Technology
* Smoothly and efficiently controlling the emulsifier process with digital frequency tracing technology of handpiece
* Easy operation and user-friendly interface with large TFT touch screen and functional menu
* Up to 10 sets of parameters preset meet different requirements and simplify the surgery operation
* Auto test checks tubing and handpiece problems (e.g.: block or leakage)
* The peristaltic pump gives complete control of aspiration flow rate as well as freedom from anterior chamber shallowing. This way can give stronger suction when working in I/A
* Built-in Vitrectomy and Cautery
* Intelligent Voice Reminder
Specifications
* Ultrasonic Emulsification (U/S)
1. Frequency: 40kHz
2. Functional mode: Continuance, Pulse, Linear, Burst
Continuance: Emulsifying power Range: 0 ~ 100%
Pulse: Frequency Range 1 ~ 99Hz
Linear: Emulsifying power linearly controlled by the footswitch, varies between 0 ~ 100%
Burst-Stopping Time: 1.5 ~ 0s
Burst-Working Time: 5 ~ 100ms
3. Maximum Vacuum: 500 mmHg
4. Maximum Flux Rate: 40 ml/min
* Irrigation/Aspiration --- Peristaltic Pump
Flux Rate: 0 ~ 40 ml/min
Reflux Rate: 20 ml/min
Max Vacuum: 0 ~ 500 mmHg
* VIT (Vitrectomy)
Cutting Power Source: Air Pump
Cutting Frequency: 20 ~ 600/min, step 10/min
Maximum Vacuum: 500 mmHg
Maximum Flux Rate: 40 ml/min
* Cautery
Maximum Power: 10 W ± 20 % (200Ω)
Type: Bipolar
Power preset between: 0 ~ 100%